Firmware Flash  ToxiPro

 
 
ToxiPro - O2, Cl2, NO2, ClO2

Version

8.30B

Posted

2008 JAN 25

Size

32 kB

Type

Binary Flash

 
 
Beta Version
 

Version

 

Posted

 

Size

 

Type

 

 

Corrects issue with saving calibration data for units that are configured to not be able to turn off.
 
 
 
Comments

To upgrade your instrument with a new firmware flash you must use the Flash Upload utility software.
The ToxiPro requires the use of IrDA communication to upload the firmware flash.

Note: CO, H2S, SO2, NH3, PH3, HCN versions of the ToxiPro uses a different firmware.  See link above.
If an incorrect firmware version is installed the instrument will display fault #7 (F 7), incorrect sensor version.  To reload the correct version simply repeat the procedure and select the correct firmware flash type

For detailed instructions on uploading a new flash into any of our instruments, click here.